OXYGEN SAVER

 

UNIQUE MODULAR INTERNAL DESIGN

» Unit designed to be switched only by the clinician between two modes of operation:

  • Average 6: 1 conversation rate in mode A.
  • Average 3:1 conversation rate in mode B.

» Pulse modes offer up to 6 lpm equivalency

Patient-selectable continuous flow feature (except Model OC006-2)

» New regular design includes brass components

» Uses one “D” size battery for greatest ambulation time

» Constant monitoring of battery level on each pulse flow, with visual and audible low battery alarms

» Audible/visual alarms serve to signal patient on system detections:

  • Cannula disconnect or kinked tubing
  • Unit inadvertently left in “on” position when not in use.

 

The volume/ time curve shown below represents typical ventilation cycle. At 20 breaths per minute and a1:2 ratio of in halation to expiration, the duration of each breath is three seconds, with one second for inhalation. Approximately .5 seconds is required for dead- space, which allows only .5 seconds as early inhalation.

Note: alveolar exchange represents only one-sixth of the ventilatory cycle.


ImPulse Elite...

Clinicians overwhelmingly agree that the preferred choice in oxygen conservation technology is every-breath pulse flow delivery. AirSep’s Impulse Elite offers two modes of every-breath delivery-depending on a patient's medical assessment.

Both delivery modes offer clear advantages to the patient. And now, Airsep, a pioneer manufacturer in the OCD technology that combines both delivery modes within one unique 12-setting unit, leaves it to you- the clinician- to Choose which method of oxygen delivery is most appropriate far your patient's requirements in maintaining oxygen saturation while providing maximum ambulation time.

Recognizing that patients' needs for supplemental oxygen are different, home care providers can rest assured that their choice of ImPulse Elite meets patients' and their clinicians' expectations.

With ImPulse Elite, an easy-to-adjust internal switch allows the provider to select an appropriate pulse mode for a particular patient. In one mode ambulation time is maximized; in the other, oxygen flow is maximized. In addition, the readily accessible flow selector provides 1-6 lpm equivalency. Economical and efficient. ImPulse Elite - the choice is yours.

Convenience and Portability.

ImPulse Elite combines convenience and portability for patients with therapeutic oxygen needs. Through advanced electronic features, the ImPulse Elite unit responds effectively to a patient's respiratory cycle and delivers the precise amount of prescribed oxygen when it is most beneficial - at the beginning of inhalation.

 

 

 

 

A versatile, self-contained system.

Realizing that the needs of patients vary greatly, AirSep, recognized worldwide for expert craftsmanship in the design and manufacture of oxygen systems, presents an oxygen conserving device - a total, self contained system that is so versatile, any ambulatory patient can use ImPulse Elite, regardless of prescription.

ImPulse Elite Ensures lightweight
portability and unobstrusive use

 

And economical.
ImPulse Elite allows you to reduce inventory by giving you the unique advantage of two OCD system technologies in one unit. So whatever the doctor orders, ImPulse Elite meets the need.

 

The unit securely and easily fits over the
top of any standard post-valve cylinder

 

Patient Friendly.

No confusing supply tubes or regulators to connect. Simple over-the-cylinder installment when empty cylinders need replacing. ImPulse Elite accepts any standard cannula and post-style cylinder. Three-light system assures patient of unit’s self-monitoring capability.

ImPulse Elite works well with E-cartable
systems, which lengthen ambulance time

Carrying Case Options:

The universal-style carrying case, specially designed for use with impulse Elite & an cylinder, serves as a vertical or horizontal shoulder bag or as a backpack. A tote-style carrying case is roomy enough for the storage of normal purse Size contents. The tote-style carrying case is used with an ML6 cylinder.

ImPulse Elite Syestem Shown with
cylinder and universal-style carrying case


S P E C I F I C A T I O N S
WEIGHT 24 OZ. (with “D” size battery) 682g

DIMENSIONS 4.39in. x 5.58in. x 2.55in.
11.15cm x 14.17cm x 6.48cm
POWER (1) 1.5V, “D” size alkaline battery
BATTERY LIFE EXPECTANCY* Mode A: 18 months Mode B: 10 months
INDICATORS Alarm/cannula disconnect
Low Battery-Audible/visual (yellow)
Pulse flow-Visual (green); indicates good battery

DELIVERY PULSE FLOW VOLUMES

Mode A
Flow Setting 1 2 3 4 5 6
mL of Oxygen 8.75 17.5 26.25 35 44 52
Actuation Frequency Every breath Every breath Every breath Every breath Every breath Every breath

Mode B
Flow Setting 1 2 3 4 5 6
mL of Oxygen 16.5 33.0 49.5 66.0 82.5 99.0
Actuation Frequency Every breath Every breath Every breath Every breath Every breath Every breath

CONSERVATION RATE MODE A: Every breath-6: 1 average
MODE B: Every breath-3: 1 average
CONTINUOUS FLOW
(Except Model OC006-2)
2 lpm oxygen; manual switch
CONTENTS GAUGE Red/Refill Zone : 0-500 psi
Green shaded Zone : 500-2,200 psi
Green Outlined Zone : 2,200-3,000 psi
MAXIMUM CYLINDER PRESSURE 3,000 psi
WARRANTY 5-year on Impulse Elite unit

 


* Based on a 2 lpm flow setting, 20 bpm, 2 hrs. per day

Home | Oxygen Therapy | Application | Technical | Products | Method of use | Dos & Donts | Press Release | Contact
Designed & Hosted : Indweb